Acoustic measurement of the triple point of neon T Ne and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry - Cnam - Conservatoire national des arts et métiers Accéder directement au contenu
Article Dans Une Revue Metrologia Année : 2021

Acoustic measurement of the triple point of neon T Ne and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry

Haiyang Zhang
Bo Gao
  • Fonction : Auteur
  • PersonId : 1139505

Résumé

We used absolute primary acoustic gas thermometry (AGT) to calibrate a Pt–Co resistance thermometer on the thermodynamic temperature scale by measuring the speed of sound in helium at a temperature T* chosen to be near the temperature of the triple point of neon, TNe. Prior to the present AGT, the Pt–Co thermometer was used with a neon triple-point cell as part of an interlaboratory comparison. Taken together, the results of the interlaboratory comparison and the present AGT redetermined the thermodynamic temperature TNe = (24.555 15 ± 0.000 24) K. This new value of TNe is consistent with other recent determinations obtained with various primary methods. After completing the AGT thermodynamic calibration, we used the Pt–Co thermometer to link T* to the temperature ratios measured by single-pressure refractive-index gas thermometry (SPRIGT) in a different laboratory. (Gao et al 2020 Metrologia 57 065006) Now, the T*-linked SPRIGT system can calibrate other thermometers on the thermodynamic temperature scale T in the range 5 K ⩽ T ⩽ TNe without using the international temperature scale ITS-90. At most temperatures in this range, the uncertainties of the T*-linked SPRIGT system are smaller than those of the ITS-90 systems used by National Metrology Institutes to calibrate resistance thermometers.
Fichier principal
Vignette du fichier
Pan_2021_Metrologia_58_045006.pdf (2.77 Mo) Télécharger le fichier
Origine : Publication financée par une institution

Dates et versions

hal-03456653 , version 1 (30-11-2021)

Licence

Paternité

Identifiants

Citer

Changzhao Pan, Fernando Sparasci, Haiyang Zhang, Pascal Gambette, Mark Plimmer, et al.. Acoustic measurement of the triple point of neon T Ne and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry. Metrologia, 2021, 58 (4), pp.045006. ⟨10.1088/1681-7575/ac0711⟩. ⟨hal-03456653⟩
49 Consultations
72 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More