Comparison of ITS-90 realizations from 13 K to 273 K between LNE-CNAM and INRIM

Abstract : Since Key Comparison CCT-K2, only few comparisons as regards realizations of cryogenic fixed points have been carried out, either bilateral (three, of which one still pending) or multilateral (two trilateral). Since a forthcoming follow-up is unlikely, any bilateral comparison, apart from key comparisons, is most welcome to provide evidence of the continuing thermometric capabilities of the laboratories involved in the realization of the ITS-90. Not too long after the establishment in 2018 of the Joint Research Laboratory for Fluid Metrology "Evangelista Torricelli" between LNE-CNAM and INRIM, the two laboratories agreed to perform a bilateral comparison, involving all the cryogenic fixed points of the scale between 13.8033 K and 273.16 Kincluding the two hydrogen vapour-pressure temperatures-, required to cover the full platinum resistancethermometer range, as dealt with in section 3.3.1 of the text underlying the ITS-90. The results are reported here and show a substantial agreement within the combined uncertainties.
Complete list of metadata

https://hal-cnam.archives-ouvertes.fr/hal-03602545
Contributor : Marie-Liesse Bertram Connect in order to contact the contributor
Submitted on : Wednesday, March 9, 2022 - 11:05:58 AM
Last modification on : Wednesday, September 28, 2022 - 5:57:02 AM

File

2007.11340.pdf
Files produced by the author(s)

Licence

Copyright

Identifiers

Collections

Citation

Dario Imbraguglio, Petrus Paulus Maria Steur, Fernando Sparasci. Comparison of ITS-90 realizations from 13 K to 273 K between LNE-CNAM and INRIM. Measurement - Journal of the International Measurement Confederation (IMEKO), Elsevier, 2020, 166, pp.108225. ⟨10.1016/j.measurement.2020.108225⟩. ⟨hal-03602545⟩

Share

Metrics

Record views

12

Files downloads

17